MAXX SP5 S - 4 x 14 L
PART NO: M-V1300414
The SP5 B compact sampler has a thermostatically controlled cooling/heating plastic housing. It is particularly suitable for high ambient temperatures.
It comes complete with a vacuum dosing system, which has a plastic metering vessel (20-350ml), motor-driven distributor tie bar and 4 x 14L container.
The sampler comes ready for operation with lead cable, shock-proof plug and a 5m suction hose.
Main Features:
- Highly accurate sample volume
- Minimal effort für calibration
- Clear operating structure and simple programming
- LAN/UMTS/GPRS Web communication (Option)
- Easy cleaning
- Modern and ergonomic design
- Big variant of bottle configuration
- Well insulated
Specification
Type | Fixed site sampler |
Housing | Made of PE with 50 mm insulation/Styrosun/PC (GF10). Housing separated in lower part (sample compartment) and upper part (control). Each part with lockable door resp. hood. Protective hood (made of Styrosun) which can be opened for operation of the control unit and service at the dosing unit. |
Thermostatic control | Self-contained, controlled cooling / heating with 4 settings, no-frost. independent of the programmable controller, Temperature in sample compartment: 4°C (adjustable from 0,0-9,9°C) |
Control | Microprocessor control, Sleep-Mode (<5mA), power supply 8-16 V foil keyboard (with keys 0-9, ESC, ENT, cursor), graphical display (128*64 Pixel), back lit |
Data logger | 3000 entries, non-volatile data memory; storage of sampling and malfunction data like sample extractions, bottle changes, messages, external signals. optional with WEB-board 100 MB (2 Year ring memory-FIFO at 1 min interval) |
Programming | 12 freely programmable user programs, with function to link programs. |
Program start options | - IMMEDIATELY; - DATE/TIME - WEEKDAY/TIME; - BY AN EXTERNAL SIGNAL |
Program End/Stop options | End of sampling program - AFTER 1 RUN - AFTER X RUNS - CONTINUOUS OPERATION - DATE/TIME |
Pause mode | Interruption of program run at any time |
Overfilling protection | Adjustable from 1–999 samples/bottle |
Interval setting | 1 min. to 99 h 59 min. in steps of 1 minute |
Pulse setting | 1 to 9999 pulses/sample |
Manual sample extraction | Possible at any time without interrupting the current program run |
Program protection | Up to 5 years after voltage loss |
Interface | Mini-USB, RS 232 optional: Ethernet RJ45, SDI-12 |
Communication (option) | Optional: Modbus, Profibus DP Connection or LAN / WLAN / GPRS-UMTS optional: 1. Connection via USB and PC |
2. LAN Modul RJ45 via TCP/IP and IE-Browser or alternatively | |
3. LAN Modul RJ45 + GPRS/UMTS Router additionaly + Fully integrated Router (industrial standard) + UMTS / GPRS + SIM card holder + E-Mail error messaging + antenna | |
Languages | Multi-language, selectable |
Signal inputs | option: expandable with 4x digital, 3 inputs can be programmed freely, and 8x analogue 0- 20 mA or 0-10 V, Impulslength 60ms, switching level 7-24 V, max. working restistance 500 Ohm, max. length of signalcable 30 m |
Signal outputs / status messages | 1x of them as collective malfunction message (Relay optional) option: expandable with 8 digital, 5 are freely programmable (in total 6 messages) |
Sampling method | -Vacuum system 20-350 ml Option: vacuum VAR flow-proportional system 5-250 ml Option: peristaltic pump 10-10.000 ml (flow-proportional) |
Single sample volume accuracy | Vacuum system: < 2,5 % or +/- 3 ml Peristaltic pump: +- 5 % or +/- 5 ml |
Suction height | Max. 7,5 m (at 1013hPa and stagnant medium), optional 8,5 m or 15 m (Power Booster) |
Pumping speed | >0,5 m/s at suction height up to 6 m (at 1013h Pa); pump capacity can be adjusted electronically |
Suction hose | PVC, L=7,5 m, ID=10 mm. Max. hose length 30 m |
Sampling modes | Time-related, flow-dependent, event-related, manual sample extraction, Option: flow-proportional (for Peristaltic Pump as standard) |
Bottle variants | Plastic 1 x 25 L 4 x 14 L 4 x 10 L 12 x 2,9 L, 24 x 1,0 L Glass 12 x 2,0 L 24 x 1,0 L |
Overall dimensions | (Hxwxd) 1.100 (1.640*) x 760 x 745 mm *) with opened top |
Weight | Approx. 75 kg with composite container, higher weight when using several bottles and/or glass bottles |
Power supply | 230 V / 115 V /AC |
Power requirement | Approx. 350VA (with cooling) |
Ambient temperature | -20 – 50°C |
Sample temperature | 0 – 40°C |
Standards | CE Sampling according to ISO 5667-10, EN16479 |
Wetted materials | PC, PVC, Silicone, PS, PE, EPDM (optional: metering vessel glass Duran50, sinker weight SS304) |